ECG Recording: Position, Electrodes, and Leads

Estimated reading time: 9 minutes

ECG strip from a dog with dilated cardiomyopathy showing two ventricular premature beats recorded in leads I, II, and III, indicated by arrows.
Kalumet (PNG conversion: Brighterorange), CC BY-SA 3.0 http://creativecommons.org/licenses/by-sa/3.0/, via Wikimedia Commons

An electrocardiogram (ECG ) is a recording of the heart's electrical activity using electrodes placed on the skin. The reliability of the recording depends on the position, the placement of the electrodes, the device settings, and the prevention of artifacts. How to read the recording is discussed in the ECG Review Guide , which includes a short section for probe holders.


1. Patient Positioning for ECG Recording

A standard ECG is taken with the animal lying on its right side (right lateral position); the vast majority of published reference values, such as those in the Reference Ranges and Evaluation document, have been obtained in this position. The head and neck are resting on the table and kept in line with the spine; the front and hind legs are held parallel to each other and perpendicular to the body. Cats tucking their legs under their bodies can lead to artificial deviations in the measurements.

The animal is placed on the floor, a rubber mat, or a non-metallic surface, not a metal table, because conductive surfaces can cause interference.

• Alternative positions : In large, uncooperative, or respiratory distressed dogs, recordings can be taken standing, lying on the chest (sternal), or lying on the left side; in cats that are fussy or have significant respiratory distress, recordings can be taken sternally or lying on the left side. Animals in respiratory distress should not be forced. Some cats sit hunched over, while others remain calm in their carrier after the electrodes are attached.

• Effect of position on recording : These positions are often sufficient for speed and rhythm, but the magnitude and direction of the waves change. In dogs filmed standing up, the Q and R waves in derivations I and II are amplified, and the electrical axis shifts to the left. In cats in the left lateral or sternal position, the R wave is usually smaller.

Sedatives are not ideally used because they can alter heart rate and rhythm. Any unusual positioning or sedation is noted.


2. Electrode Placement and Color Codes

Electrodes are attached to the skin with alligator clips, metal plates, or adhesive pads. Clips that can be applied without cutting the fur are the most common; their ends should be flattened so they don't pinch the animal. Gel pads that attach to paw pads are more comfortable and better tolerated by cats.

• Front legs : The clip is placed just above the elbow or at the tip of the elbow (olecranon); the presence of less muscle underneath reduces the parasite. Since a clip too close to the chest may move with respiration, it can also be used between the elbow and wrist.

• Hind legs : The clip is attached to the skin above the knee joint, above the patellar tendon.

Extremity electrodes are identified by a standard color code, and there are two different systems:

• European system (most countries) : right front leg red, left front leg yellow, left hind leg green, right hind leg black.

• American system : right front leg white, left front leg black, left hind leg red, right hind leg green.

The right rear leg electrode is the grounding electrode. When the two front leg electrodes are swapped, the P wave and QRS complex appear negative in derivation I. Since the P wave is almost always positive in derivations I, II, and III, the wiring should be checked first when a negative P wave is observed.

To improve contact, the skin is moistened with a conductive liquid or ECG gel. Clips often use alcohol; however, since alcohol is flammable and evaporates in 5–10 minutes, gel may be preferred for longer recordings. Gel pads do not require alcohol. If the alcohol from one electrode transfers to another, all leads will look similar; electrodes should not touch each other, the person holding them, or the table, and the wires should not cross.


3. Extremity and Chest Derivations

The derivation is a recording of the potential difference between a positive and a negative electrode, offering a different perspective on the heart. The derivation is upward if the activity is moving towards the positive electrode, and downward if it is moving away. The propagation of the stimulus is described in the article "Electrical Activity of the Heart".

• Bipolar (standard) derivations : The two anterior limbs and the left hind limb form Einthoven's triangle. Derivation I records the difference between the right anterior (negative) and the left anterior limb (positive), derivation II records the difference between the right anterior (negative) and the left hind limb (positive), and derivation III records the difference between the left anterior (negative) and the left hind limb (positive).

• Unipolar (amplified) derivations : In aVR, the right anterior leg is the positive electrode; in aVL, the left anterior leg; and in aVF, the left posterior leg is the positive electrode; the average of the other two legs becomes the neutral reference. Since deviations are smaller, they are amplified by the device.

These six derivations form a six-axis system representing the frontal plane. The calculation of the electrical axis is explained in the article "Electrical Axis of the Heart (MEA)".

Chest (precordial) leads are unipolar leads that represent electrical activity in the transverse plane. The Wilson system is most commonly used in veterinary medicine; in dogs, the placement is as follows:

• V1 : In the space between the right 5th ribs, just next to the sternum.

• V2 : In the space between the left 6th ribs, just next to the sternum.

• V4 : In the space between the left 6th ribs, where it joins the rib cartilage.

• V3 : Right in the middle of V2 and V4.

• V5 and V6 : In the same space, above V4 towards the back, spaced sequentially at intervals equal to the distance between V3 and V4.

Chest leads are useful in supporting the diagnosis of chamber enlargement and intraventricular conduction disorders, and in finding elusive P waves. However, because chest structure is highly variable in dogs, applying a single system to every animal is difficult; they are not necessary in most recordings.


4. Paper Speed, Calibration, and Filters

• Paper speed : Most commonly used is 25 or 50 mm/s; 25 mm/s is preferred for normal heart rates in dogs, and 50 mm/s is preferred for rapid heart rate and in cats. Calculating heart rate is explained in the article "Determining Heart Rate and Rhythm".

• Calibration : The standard setting is 10 mm/mV. If the complexes are very small, especially in cats, 20 mm/mV is selected; if they are very large or the derivations overlap, 5 mm/mV is chosen. The rectangular calibration mark at the beginning of the recording is usually 1 mV high and 0,2 seconds wide; the rate and calibration must be checked before reading the recording.

• Filters : Low-pass filters filter signals above a certain frequency, while high-pass filters filter signals below it; notch filters target 50–60 Hz interference from the network. Each filter distorts the signal and can particularly reduce the R-wave. If filters are to be used, the low-pass filter should be set to 50–70 Hz for dogs and 150 Hz for cats; the high-pass filter should be set to 0,05 Hz.

If there is good contact, a better trace is obtained with the filter off in most recordings, and wave heights are measured in the unfiltered recording.


5. Artifacts and How to Prevent Them?

Artifacts are deviations that do not originate from the heart and can mask true waves or mimic an arrhythmia.

• Muscle tremor : Tremors, standing, and purring cause irregular, subtle undulations in the baseline and can mask P waves. Keeping the animal calm, preferably in a comfortable position rather than standing, and supporting its legs helps. When the cat purrs, it's recommended not to pet it, but to turn on a tap or place an alcohol-soaked cotton ball near its face.

• Breathing and movement : Respiratory movements and rapid mouth breathing (panting) undulate the baseline, and leg movement creates sudden shifts. Placing one hand on the chest wall can reduce the respiratory effect.

• Electrical interference : These are subtle, regular fluctuations repeating 50 (60 in America) times per second and are often caused by poor clip-to-skin contact. Improving contact can be helped by placing an insulating mat under the animal, grounding the device, and turning off other electrical appliances in the room.

• Poor electrode contact : The baseline thickens, R waves become smaller, and the recording may become unreadable.

If the parasite is in derivations I and II, the right anterior leg electrode is checked; if in I and III, the left anterior leg electrode is checked; and if in II and III, the left hind leg electrode is checked. In multi-channel recording, an unaffected derivation reveals a motion artifact mimicking a premature beat.


6. Recording Time, Rhythm Tape and Holter

As many leads as possible are recorded: at least 3–5 minutes if 12 leads are recorded simultaneously, or at least 10 seconds each if recorded sequentially. Then, a long rhythm strip is usually taken from lead II and continued until intermittent abnormal beats heard on auscultation are detected. Position, velocity, calibration, and filter are noted in the recording.

• Smartphone devices : The electrode sheath is pressed against the left chest wall. In 46 dogs and 23 cats, the instantaneous heart rate was the same as with standard ECG, and the 15-second rhythm recordings were found to be very accurate; however, the direction of the complexes differed occasionally in dogs and frequently in cats. These devices are a tool that complements standard ECG.

A clinical ECG, lasting approximately 2 minutes, records only about 0,1% of a dog's daily heartbeats. In 35,8% of studies of affected Dobermann Pinschers, no early ventricular beats were observed on a 5-minute ECG. Holter monitoring , on the other hand, records the ECG continuously for 24 or 48 hours. Its main applications are as follows:

• Intermittent symptoms : investigation to determine if fainting or weakness is caused by an arrhythmia ( Syncope (Fainting), Seizure and Collapse) ).

• A known rhythm disorder : determining its extent and severity.

• Treatment evaluation : monitoring the need for and effectiveness of rhythm therapy; assessing heart rate control in atrial fibrillation.

• Breed screening : Dilated cardiomyopathy screening in Dobermann Pinschers , and arrhythmogenic right ventricular cardiomyopathy (ARVC) screening in Boxers.


7. For Owners: What Happens During an ECG?

An ECG is a recording of the heart's electrical signals reaching the skin's surface; nothing is inserted into the body. Your cat or dog is usually gently held on its right side, and small clips or adhesive pads are attached to its legs. The clips are usually applied without trimming the fur; the skin is moistened with alcohol or gel.

The recording takes a few minutes and can be done during the examination, even in the owner's presence. It's important that the animal remains calm, as trembling, purring, and movement can disrupt the recording. Sedatives are avoided as much as possible; animals experiencing respiratory distress are not forced to lie down. If necessary, a small area of ​​fur is trimmed, and the animal wears the device for 24 or 48 hours.


8. The Role of ECG in Cardiology Examination

ECG is a primary tool for detecting rhythm and conduction disorders; however, its sensitivity in detecting chamber enlargement is low, meaning a normal ECG does not rule out enlargement. Irregular rhythm heard on auscultation, fainting, weakness, or exercise intolerance are the main causes of ECG abnormalities.

A cardiology examination proceeds in the following order:

• Information form : history, complaints, past illnesses, and respiratory rate.

• Physical examination : including auscultation.

• Chest X-ray in two positions

• Echocardiography

• ECG : assessment of rhythm and conduction; Holter monitoring if necessary.

• Blood pressure measurement

• Blood tests : determined if deemed necessary.

Every cat and dog, regardless of age, should undergo a cardiology examination before receiving anesthesia; an ECG is also frequently monitored during anesthesia. For details, please see the article on Pre-Anesthesia Cardiology Examination .

Cats and dogs exhibiting irregular heartbeat, fainting, or lethargy should have their hearts evaluated by a veterinary cardiologist.


How Did You Find the Article?
Your evaluations in this section will be taken into consideration but will not be published.
Did you find the content useful?


Home » ECG Recording: Position, Electrodes, and Leads


Other Articles You May Be Interested In


"It is impossible to separate naming from science, and science from naming."

Antoine Lavoisier

Copyright © 2026 Veterinary Cardiology Portal. All rights reserved. Copyright and Terms of Use